South Australian Consolidated Regulations89—Apparatus used for industrial radiography
(1) Apparatus used for
industrial radiography must be durably marked with a label containing the
following information:
(a) the
serial number of the apparatus; and
(b) the
maximum accelerating voltage (kV) at which the tube can be operated; and
(c) the
maximum continuous current (mA) at which the tube can be operated.
(2) Apparatus used for
industrial radiography must—
(a) have
a key operated device which controls, and when locked, prevents the supply of
power to the high voltage generator; and
(b) have
a red or amber fail safe light that indicates when X-rays are being produced.