South Australian Consolidated RegulationsSchedule 9—Error distances for automatic collimation to an image
intensifier
1 For the purposes of this
Schedule—
area being imaged means the area of the input phosphor which produces an image
on the television monitor;
error distance means the lack of alignment between the X-ray field and the
area being imaged, where the X-ray field lies outside the area being imaged.
2 For a polygonal X-ray field, measurements of the
error distance are taken perpendicularly from the mid-point of each side of
the X-ray field, which is outside of the area being imaged, to the
corresponding boundary of the area being imaged.
3 For an X-ray field with a curved boundary (eg a
circular X-ray field) the error distance is defined for all points on the
boundary of the X-ray field which lie outside of the area being imaged. For
any such point the error distance is measured perpendicularly from the tangent
to the boundary at that point to the corresponding boundary of the area being
imaged.
4 In no case must the error distance, measured in
the way described above, exceed 1 per cent of the focal spot to image receptor
distance.